Print a range of records, Print all records, Print records by type – Fluke Biomedical medTester 5000C User Manual
Page 113: Conducting waveform tests in safety autosequences

Biomedical Test System
Performance Waves
7
7-5
Conducting Waveform Tests in Safety
Autosequences
You can execute performance wave tests at the end of any safety
autosequence. The medTester 5000C prompts you to begin or to skip these
tests.
When performance waveforms are generated within a safety autosequence,
they follow this sequence for the listed periods of time:
Performance Waveforms in Safety Autosequences
Waveform Name
Test Duration (In Seconds)
2 Hz Square
Continuous
4 sec Step
8
10 Hz Sine
4
40 Hz Sine
4
60 Hz Sine
60
100 Hz Sine
4
1 kHz Square
4
2 Hz Triangle
Continuous
CMRR 60 Hz
Continuous
ECG 60 BPM
Continuous
ECG 30 BPM
30
ECG 60 BPM
30
ECG 120 BPM
30
ECG 240 BPM
30
Note
Tests with a predefined duration automatically end or can be
terminated at any time by pressing
(NEXT)
. Those marked
continuous can be terminated by pressing
NEXT
.
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)