Downloading – Fluke Biomedical medTester 5000C User Manual
Page 185

Biomedical Test System
Remote Operation
12
12-7
Forcing A Port
An alternative to the above procedure is forcing a port. For example, if you
want COM1 to be the remote port, type or send
remote1
. This will force
COM1 to act as the remote port. To force COM2, type or send
remote2
.
If you have made a remote connection, the medTester responds by sending:
REMOTE MODE
out the remote serial port. At the same time, the
medTester 5000C display becomes blank.
Command Syntax Rules
•
Send commands in uppercase or lowercase letters.
•
Spaces are ignored.
•
Use the backspace character for editing.
•
Terminate commands with a carriage return (CR), or line feed (LF),
or both.
•
Commands with an equal sign (=) require that at least one parameter
be included with the command.
•
Use commas to separate multiple parameters.
Complete lists of medTester 5000C remote commands and parameters are
shown in the Remote Commands List section later in this chapter.
Responses to Commands
•
Every command you issue to the medTester returns a response to
the remote serial port.
•
Responses return character strings, numbers, or data files.
•
If no other response appears, an asterisk (*) appears to indicate that
the command has been executed.
•
Previous commands must execute and return their responses before
the medTester will accept further commands.
•
The null command (consisting only of terminating characters)
returns a question mark (?).
•
Command responses are in uppercase letters followed by a carriage
return and line feed.
•
Measurement reading responses appear as they do on the
medTester 5000C. Numbers may appear with or without a decimal
point and are followed by the unit of measurement abbreviation.
See the table that follows.