For the index 2xl – Fluke Biomedical medTester 5000C User Manual
Page 248

medTester 5000C 
Operators Manual 
15-20
For the Index 2XL:
SPO2 Autosequence Number: A72
Autosequence Name: PMS
Manufacturer: PMS M1190
Settling Time: 25 Beats
Heart Rate Limit:
± 4 BPM (beats per minute)
Step Number
SPO2
(in percent
SPO2 Limit
(in percent)
Heart Rate
(in BPM)
Amplitude
(in percent)
Base TLC
1 96
± 3
75 6.0 3500
2 98
± 3
230 6.0 3500
3 92
± 3
75 6.0 3500
4 82
± 4
75 6.0 3500
5 72
± 5
75 6.0 3500
6 98
± 3
45 6.0 3500
Starting Alarm Value for SPO2: 96% 
Starting Alarm Value for Rate: 75 BPM 
Starting Alarm Value for Amplitude: 6.0% 
SPO2 Alarm Test: Hi–100% Lo–85% 
Rate Alarm Test: Hi–140 BPM Lo–50 BPM 
Asystole Test 
Amplitude Test 
Probe Test 
SPO2 Autosequence Number: A73
Autosequence Name: DATEX
Manufacturer: Datex
Settling Time: 25 Beats
Heart Rate Limit:
± 4 BPM (beats per minute)
Step Number
SPO2
(in percent
SPO2 Limit
(in percent)
Heart Rate
(in BPM)
Amplitude
(in percent)
Base TLC
1 96
± 3
75 6.0 3030
2 98
± 3
230 6.0 3030
3 92
± 3
75 6.0 3030
4 82
± 4
75 6.0 3030
5 72
± 5
75 6.0 3030
6 98
± 3
45 6.0 3030
Starting Alarm Value for SPO2: 96% 
Starting Alarm Value for Rate: 75 BPM 
Starting Alarm Value for Amplitude: 6.0% 
SPO2 Alarm Test: Hi–100% Lo–85% 
Rate Alarm Test: Hi–140 BPM Lo–50 BPM 
Asystole Test 
Amplitude Test 
Probe Test 