Print your autosequence, User output names, Reinitialize factory default settings – Fluke Biomedical medTester 5000C User Manual
Page 224: Comments, Next test due date, User time, User time is required, Print test tag, Store or print test record

medTester 5000C
Operators Manual
14-14
Autosequence Number: A64
Autosequence Name: VLSSE4
Step and Test
Type
Load
ESU Mode
Output)
Power
Setting
Expected Power
or
Electrode
Limits
1
Generator Output
300 Ohms
Monocoag
Spray
25
25 W
± 20%
2
Generator Output
300 Ohms
Monocoag
Spray
120
120 W
± 15%
3
Generator Output
300 Ohms
Monocoag
Soft
60
60 W
± 15%
4
Generator Output
300 Ohms
Monopolar
Pure Cut
75
75 W
± 15%
5
Generator Output
300 Ohms
Monopolar
Pure Cut
300
300 W
± 15%
6
Generator Output
300 Ohms
Monopolar
Blend 1
200
200 W
± 15%
7
Generator Output
300 Ohms
Monopolar
Blend 2
200
200 W
± 15%
8
Generator Output
300 Ohms
Monopolar
Blend 3
200
200 W
± 15%
9
Generator Output
300 Ohms
Monopolar
Blend 1
75
75 W
± 15%
10
Generator Output
300 Ohms
Monopolar
Blend 2
75
75 W
± 15%
11
Generator Output
300 Ohms
Monopolar
Blend 3
75
75 W
± 15%
12
Generator Output
100 Ohms
Bicoag
Standard
70
70 W
± 15%
13
Generator Output
100 Ohms
Bicoag
Precise
70
70 W
± 15%
Posttest Prompts
When all tests are completed, the medTester 5000C prompts you for:
1.
COMMENTS
2.
NEXT TEST DUE DATE
3.
USER TIME
EMS/CMMS system users:
User time is required
4.
PRINT TEST TAG
5.
STORE or PRINT TEST RECORD