Customizing bar code record data entry, Turning prompts on and off – Fluke Biomedical medTester 5000C User Manual
Page 101

Biomedical Test System
Customize Your medTester
6
6-9
medTester Autosequence Prompts
Prompt #
Prompt Name/
Renameable?
(Yes or No)
Description
14
NEXT TEST DUE
DATE (No)
Enter the next test due date. Maximum of nine characters.
15
USER TIME (No)
Enter the time used by the user to perform the tests in hours.
The format can be six (6) digits, including a decimal point.
16 ELAPSED
TEST
TIME (No)
If the prompt is turned off, the medTester does not store the
elapsed test time with the test record. In addition, the elapsed
test time will not display or print.
If the prompt is turned on, the medTester measures the
elapsed test time for the entire autosequence and stores it
with the test record. The time will then display and print with
the record.
17
PRINT TEST TAG
(No)
This message prompts the user to print a test tag.
18 STORE
RECORD
(No)
After the autosequence is finished, the user is prompted to
store the test record to memory and/or to print the record. If
this prompt is turned off, the record is stored automatically in
memory.
Turning Prompts On and Off
You can switch any autosequence prompt on or off. Whenever you switch off
a prompt, that prompt will not appear in any autosequence. To customize
these prompts, go to the prompts menu:
PROMPTS
F3
CUSTOM
F3
The medTester 5000C displays the message:
and will next display the first prompt, its on or off status, and menu items:
CONFIGURE AUTOSEQUENCE PROMPTS
.