End-of-test prompts, Tag print function, Autosequence steps – Fluke Biomedical medTester 5000C User Manual
Page 77
Biomedical Test System
Autosequences
5
5-7
Autosequence Steps
If you have autosequence prompts turned on, you enter the appropriate
information. After responding to prompts, you see this message appear on
the medTester 5000C display:
This message is telling you to:
1.
Plug the EUT into the test receptacle.
2.
Attach a single Kelvin cable to the pair of
red
binding
posts on the patient lead panel.
3.
Attach the Kelvin cable clamp to an exposed grounded
surface on the EUT.
4.
If the autosequence involves leads, attach the leads to the
top-panel, color-coded ECG binding posts.
5.
Press
READY
F5
after the connections are made. The
sequence tests begin. Descriptions of tests follow.
Note
You can choose which tests your autosequences execute.
However, the factory default settings for all autosequences
include all of the tests described here. Chapter 6, Customize Your
medTester 5000C, shows you how to customize autosequence
tests.
PLUG IN EUT, ATTACH KELVIN CABLE AND
PATIENT
LEADS