For a pca test, Ipt-1 – Fluke Biomedical medTester 5000C User Manual
Page 158

medTester 5000C
Operators Manual
10-20
IPT-1 AUTOSEQUENCE TESTS
Nominal
Flow Rate
(ml per
hour)
Test Type
High
Limit
(ml per
hour)
Low
Limit
(ml per
hour)
End Type
(manual,
volume,
or timed)
Ending Quantity
If Applicable
(ml for volume or
hours:minutes
for timed)
Autosequence Name (Test Type): LINEAR
Keyboard Shortcut Command: A50
60.0 Steady
Flow
66.0 54.0 Volume
5.0
120.0 Steady
Flow
132.0 108.0 Volume
10.0
240 Steady
Flow
264.0 216.0 Volume
20.0
Autosequence Name (Test Type): PULSED
Keyboard Shortcut Command: A51
60.0 High
Accuracy
66.0 54.0 Volume
5.0
120.0 High
Accuracy
132.0 108.0 Volume
10.0
120.0 High
Accuracy
132.0 108.0 Volume
10.0
Autosequence Name (Test Type): LOW
Keyboard Shortcut Command: A52
10.0 Steady
Flow
11.0 9.0 Timed
00:05
Autosequence Name (Test Type): HIGH
Keyboard Shortcut Command: A53
480.0 High
Accuracy
504.0 456.0 Timed
00:05
Autosequence Name (Test Type): EVAL
Keyboard Shortcut Command: A54
5.0 Steady
Flow
5.9 4.1 Volume
1.5
10.0 Steady
Flow
11.0 9.0 Volume
2.5
20.0 High
Accuracy
22.0 18.0 Volume
2.5
30.0 High
Accuracy
33.0 27.0 Volume
2.5
60.0 High
Accuracy
66.0 54.0 Volume
5.0
120.0 High
Accuracy
132.0 108.0 Volume
10.0
240.0 High
Accuracy
264.0 216.0 Volume
30.0
480.0 High
Accuracy
528.0 432.0 Volume
60.0