Lead to lead leakage – Fluke Biomedical medTester 5000C User Manual
Page 48

medTester 5000C
Operators Manual
4-2
Hot to Neutral
Enter menu commands:
MANUAL
F2
VOLTS
F1
HOT-NEUT
F1
Neutral to Ground
Enter menu commands:
MANUAL
F2
VOLTS
F1
NEUT-GND
F2
Hot to Ground
Enter menu commands:
HOT-GND
F3
MANUAL
F2
VOLTS
F1
Leakage Current
All leakage current measurements are taken through a 1–k
Ω AAMI load RMS
or DC. Current is measured in a range of 2000
µA or 200 µA.
Note
These tests are described using Kelvin cables for connections.
However, single-wire test cables can be used. Connect single-
wire test cables to the external input posts only.
Note
For all leakage current measurements, except ECG lead
isolation,
DC
toggles the measurement between RMS and DC.
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)