For cardiosat 100 – Fluke Biomedical medTester 5000C User Manual
Page 240

medTester 5000C
Operators Manual
15-12
Probe Test
For the Oxitest PLUS, there is no probe test built-in. However, it is still
possible to include a probe test in an autosequence. In this case, simply test
the probe some other way and record the test result during the autosequence
by pressing PASS or FAIL.
The CardioSat 100 version E and EF have a built-in probe test feature
Note:
For the Cardiosat 100, if you used the electrical simulation
technique to perform the earlier tests, either disconnect the
electrical simulation adapter from the CardioSat 100 or turn the
SPO2 monitor power off before performing the probe test. This
electrical signal can interfere with the proper operation of the
probe test.
The autosequence performs the regular CardioSat 100 probe test and records
the results. Prepare for the test by connecting the probe to be tested to the
appropriate connector on the CardioSat 100 with the applicable adapter cable.
Then press
F5 START
and wait for the results.
When the test is done, you can read the results with the
DOWN/UP
keys, then
press
F5 END
. You can repeat the test if desired.
The table beginning on the next page supplies specific information about the
tests for each factory initialized autosequence.