Ground Fault Circuit Interrupter
The test receptacle is protected by a ground fault circuit interrupter. In the 
event of a fault in the EUT (when the test receptacle is powered on) which 
causes a current imbalance in the hot and neutral lines of greater than 5 mA, 
the interrupter triggers. Under this condition, the interrupter turns off the test 
receptacle and all test connections, and the medTester prompts you to correct 
the fault and continue. 
Performance Waveforms 
The medTester 5000C generates ECG waves and arrhythmias to test the 
performance of patient monitors in those medTesters that have the 
Waveform/Extended Testing Module installed. You can find information 
about conducting waveform tests in Chapter 7, Performance Waves. 
Specifications for waveforms and arrhythmias are: 
ECG Performance Test Waves (Lead I, Vp-p)
•
Square Wave—2 Hz, 1 mV
•
DC Pulse—4 Seconds, 1 mV
•
Sine Wave—0.5, 10, 40, 60, and 100 Hz, 1 mV
•
Square Wave—1 kHz, 1 mV
•
Triangle—2 Hz, 1 mV
•
CMRR—60-Hz sine wave with 1–k
Ω imbalance in LA
•
Normal Sinus—30, 60, 120, 240 BPM.
Arrhythmias
•
Atrial fibrillation
•
Second degree A-V Block, Type 1
•
Premature atrial contractions
•
Missed beat at 80 and 120 BPM
•
PVC 1 left
•
PVC 2 right
•
Multifocal PVCs
•
PVC 1, R on T
•
A pair of PVCs
•
Run of 5 PVCs
•
Run of 11 PVCs, MF
•
Right bundle branch block
•
Ventricular tachycardia
•
Ventricular fibrillation
•
Asystole