Infutest, For a single rate test, For a dual rate test – Fluke Biomedical medTester 5000C User Manual
Page 157

Biomedical Test System
IV Pump Module
10
10-19
INFUTEST 2000 Series D AUTOSEQUENCE TESTS
Nominal
Flow Rate
(ml per
hour)
Test Type
High
Limit
(ml per
hour)
Low
Limit
(ml per
hour)
End Type
(manual,
volume,
or timed)
Ending Quantity
If Applicable
(ml for volume or
hours:minutes
for timed)
Autosequence Name (Test Type): PUMP 1
Keyboard Shortcut Command: A50
60.0 Single
Rate
66.0 54.0 Volume
5.0
120.0 Single
Rate
132.0 108.0 Volume
10.0
240 Single
Rate
264.0 216.0 Volume
20.0
Autosequence Name (Test Type): PUMP 2
Keyboard Shortcut Command: A51
30.0 Single
Rate
35.0 25.0 Volume
3.0
60.0 Single
Rate
66.0 54.0 Volume
5.0
120.0 Single
Rate
132.0 108.0 Volume
10.0
240.0 Single
Test
264.0 216.0 Volume
20.0
Autosequence Name (Test Type): DUAL
Keyboard Shortcut Command: A52
Secondary
120.0
Primary
20.0
Dual Rate
132.0
25.0
108.0
15.0
Manual
Autosequence Name (Test Type): PCA
Keyboard Shortcut Command: A53
Bolus Volume
1.0ml
Lockout Time
30.00(mm:ss)
PCA 1.30ml .70ml Timed
03.00
Autosequence Name (Test Type): EVAL
Keyboard Shortcut Command: A54
5.0 Single
Rate
5.9 4.1 Volume
1.5
10.0 Single
Rate
11.0 9.0 Volume
2.5
20.0 Single
Rate
22.0 18.0 Volume
2.5
30.0 Single
Rate
33.0 27.0 Volume
2.5
60.0 Single
Rate
66.0 54.0 Volume
5.0
120.0 Single
Rate
132.0 108.0 Volume
10.0
240.0 Single
Rate
264.0 216.0 Volume
30.0
480.0 Single
Rate
528.0 432.0 Volume
60.0