Pressure test, Running nibp autosequence tests – Fluke Biomedical medTester 5000C User Manual
Page 278

medTester 5000C 
Operators Manual 
17-4
Running NIBP Autosequence Tests
To begin running NIBP autosequence tests, connect your medTester 5000C 
to your CuffLink. These two instruments are connected through the COM2 
or designated wedge port on the medTester and the RS-232 serial port on the 
CuffLink. 
Note
Refer to Chapter 1, General Information - Accessories, or the current 
Fluke Biomedical Price List for the correct serial cable part 
number needed. 
From the medTester 5000C NIBP menu, select the autosequence name for 
the autosequence you wish to run. The name and number of the selected 
autosequence appears on the display followed by the prompt: 
If a safety autosequence has executed since power up of the medTester, you 
can have the medTester use the previously input prompt information for the 
current pacer autosequence, from this prompt: 
USE PREVIOUS STARTING INFORMATION?
YES
F1
NO
F2
To accept previously input prompt information, press
YES
, and go to Test
Sequence, which follows this section. If you choose
NO
, go to the prompt
instructions in Pretest Device Prompts.
Press F5 for all device info prompts