Fluke Biomedical medTester 5000C User Manual
Page 18

medTester 5000C 
Operators Manual 
1-6
Prompts in the autosequence ask you for information that aids in test 
reporting. The pretest prompts, for example, can ask you for the model and 
serial number of the equipment you’re testing, the test location, operator 
code, and physical condition of the EUT (equipment under test). Posttest 
prompts can ask you to schedule the next test date, input comments about the 
tests, or prepare a test tag. 
With the Memory Module and RS-232/Printer Module installed on your 
medTester 5000C, you can store test records to memory and print them from 
the medTester’s printer port. Additional autosequence modules are available 
for performance testing defibrillators, IV pumps, and more. See Chapter 5, 
Autosequences. 
Computer Control 
You can remotely control your medTester 5000C from a personal computer 
or compatible terminal device through a serial communication port and 
appropriate communication software. Once connected and in remote control of 
the medTester 5000C, you send remote commands to the medTester 5000C 
to do safety and performance tests, with the results returned to the computer. 
Some medTester 5000C users prefer to accumulate the test data for 
equipment before reporting stored records. Then at their convenience they 
upload the data to an equipment database program, Computerized 
Maintenance Management System, or Equipment Management System (EMS) 
on a computer. This is accomplished through a serial port on the 
medTester 5000C and the medTester 5000C’s Data Transfer Module. Chapter 
12, Remote Operation, discusses the methods you can use to control the 
medTester 5000C and to manage test data. 
If you use other Fluke Biomedical analyzers, such as the Impulse 4000 or IPT-
1, you can control them with the medTester 5000C. By connecting an 
analyzer to the medTester 5000C’s COM2 port, you gain a two-way 
communication link. This link lets the medTester 5000C control the 
operation of the analyzer. Test data that the analyzer collects is then returned 
to storage in the medTester 5000C. 
medCheck 
The medTester 5000C medCheck Module unleashes the power of the 
medTester 5000C to become the centerpiece of a total equipment 
management solution. medCheck lets you download checklists from, and 
upload checklist test data to, your equipment management software system. 