Resetting autosequences to default settings, Printing your medtester 5000c configuration, Customize medtester 5000c options – Fluke Biomedical medTester 5000C User Manual
Page 107

Biomedical Test System
Customize Your medTester
6
6-15
Resetting Autosequences to Default Settings
You can always return medTester 5000C autosequences to the default settings
by reinitializing the medTester. Safety autosequences require you to indicate
whether you want to include power-off tests, reverse polarity tests, and open
neutral tests. To access this feature for the safety autosequences, press the
right arrow on the custom menu display, and enter the menu commands:
SAFETY
F1
INIT
F3
To return to the default settings for other types of autosequences, such as
defibrillation or IV pump, select the desired type from the custom
initialization menu. To initialize all custom features to default, press the right
arrow from the custom initialization menu, then press:
OTHER
F4
Pressing
other
sets all customized features to the factory default settings. For
more information about these settings, see Factory Default Settings in Chapter 2,
Installation.
Printing Your medTester 5000C Configuration
With a printer connected to the printer port on the back panel of your
medTester 5000C, you can print out your configuration of the medTester by
entering these commands:
F4
Customize medTester 5000C Options
The descriptions of autosequence customizing features in this chapter apply
as well to other autosequences. If your medTester 5000C includes the
defibrillator and/or IV pump modules, you can customize those
autosequences tool. For more information see:
•
Chapter 9, Defibrillator Module.
•
Chapter 10, IV Pump Module.
•
Chapter 14, ESU Module.
•
Chapter 15, SPO2 Module.