General specifications, Operating: 15( to 35(c (59( to 95(f), Ecg connections – Fluke Biomedical medTester 5000C User Manual
Page 24: Ecg binding posts, High-level ecg output, Data input and output, Rs-232c serial interface parameters, Real time clock, Medtester 5000c instrument specifications

medTester 5000C 
Operators Manual 
1-12
medTester 5000C Instrument Specifications
Line Voltage and Measurements
•
Hot to Neutral.
•
Neutral to Ground.
•
Hot to Ground.
•
Range: 200.0 V RMS.
•
Accuracy: ±5% of range.
Leakage Current 
Leakage current is measured through a 1 k
Ω AAMI load RMS or DC
measured and displayed in microamperes. These tests can be thought of as 
measuring the voltage, in millivolts across the same load. 
Leakage Current Measurements
•
Case external.
•
Case internal.
•
External.
•
ECG—Leads are RL, RA, LA, LL, and V1-V6 tied together:
1.
To ground, all leads or individual leads.
2.
Lead to lead, individual leads to all other leads together.
3.
Lead isolation (RMS only), all leads or individual leads. 
This test is with line voltage applied from the lead(s) to 
ground, current limited by 120 k
Ω.
Note
If you receive abnormally high leakage current measurements 
with V1-V6 leads tied together, you can inspect individual V 
leads by removing the other leads. This procedure allows you to 
isolate the lead with a high leakage current reading. 