Fluke Biomedical medTester 5000C User Manual
Page 254
medTester 5000C
Operators Manual
15-26
MORE
F4
to select more steps, or
END
F5
to end your selections.
6.
STARTING ALARM VALUES: Enter starting values for the alarm
tests for:
•
SPO2
•
HEART RATE
•
AMPLITUDE (For the Index 2XL, enter amplitude from 0% to
20% with the thirty-three selections offered)
Note
Even if you do not intend to perform any alarm tests, you still
have to go through the starting alarm values. You can just press
ENT
to get through them.
7.
Answer F4 YES or F5 NO whether to perform the following special tests:
•
SPO2 ALARM TEST: If YES, then enter the Hi and Lo alarm
values.
•
RATE ALARM TEST: If YES, then enter the Hi and Lo alarm
values.
•
ASYSTOLE TEST
•
AMPLITUDE TEST
•
MOTION TEST (Index 2XL only) – Select the motion test desired.
•
PROBE TEST
After customizing your autosequence, the medTester prompts you to store
the autosequence:
STORE
F5
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)