Medcheck module, Tap your medtester™s potential – Fluke Biomedical medTester 5000C User Manual
Page 162

medTester 5000C
Operators Manual
10-24
2.
SECONDARY RATE HIGH LIMIT: Enter the high limit for the
flow rate. 0000.0 turns off high limit checking.
3.
SECONDARY RATE LOW LIMIT: Enter the low limit for the
flow rate. 0000.0 turns off low limit checking.
4.
PRIMARY FLOW RATE: Enter the nominal flow rate, 0000.1 to
0999.0 ml/hr.
5.
PRIMARY RATE HIGH LIMIT: Enter the high limit for the
flow rate. 0000.0 turns off high limit checking.
6.
PRIMARY RATE LOW LIMIT: Enter the low limit for the flow
rate. 0000.0 turns off low limit checking.
7.
TEST END TYPE: Select
8.
MANUAL to end the test by pressing a key.
9.
TIMED to end the test after a specific time:
10.
END TIME: Enter the elapsed time that will end the
test, 00:01 to 99:59 (hours:minutes).
11.
VOLUME to end the test after a specific volume:
12.
END VOLUME: Enter the volume that will end the
test, 0000.1 to 9999.0 ml.
For a PCA Test:
• BOLUS VOLUME: Enter the bolus volume, 00.50 to 99.99 ml.
• BOLUS VOLUME HIGH LIMIT: Enter the high limit for the
bolus volume. 00.00 turns off high limit checking.
• BOLUS VOLUME LOW LIMIT: Enter the low limit for the
bolus volume. 00.00 turns off low limit checking.
• LOCKOUT TIME: Enter the lockout time between boluses, 00:01
to 99:99 (minutes:seconds).
• END TIME: Enter the elapsed time that will end the test, 00:01
to 99:99 (hours:minutes).
IPT-1
For the IPT-1, there are 2 types of tests available, steady flow and high
accuracy:
1.
FLOW RATE: Enter the nominal flow rate, 0000.1 to 0999.0
ml/hr.
2.
FLOW TEST TYPE: Enter STEADY for steady flow or HI ACC
for high accuracy.
3.
RATE HIGH LIMIT: Enter the high limit for the flow rate.
0000.0 turns off high limit checking.