Reinitialize factory default settings, Customizing iv pump tests – Fluke Biomedical medTester 5000C User Manual
Page 160

medTester 5000C 
Operators Manual 
10-22
Customizing IV Pump Tests
 
(For detailed information about customizing autosequences, please refer to 
Chapter 6, Customize Your medTester 5000C.) 
To begin customizing IV pump autosequences, you can view any of your 
current ones by entering these menu commands: 
1.
From the main menu,
MENU1
, press the right arrow to access
MENU2.
2.
Enter the menu commands:
IVPUMP
F3
AUTOSEQ
F1
CUSTOM
F3
3.
From the
IVPUMP
menu, select:
VIEW
F2
4.
Now select the IV pump autosequence you want to view.
5.
Finally scroll through the steps in the sequence by using your up 
and down arrow keys. 
Make Your Own Autosequence 
You can customize the five preprogrammed IV pump autosequences or 
create five new ones with the 
blank
sequences. This section describes the
menu steps and order of steps that you use to customize, or create, 
autosequences. Note that each customized autosequence can include more 
than one step. After you complete item one below (name), the 
medTester 5000C prompts you for step one entries which begin with the flow 
tests described in number two below. You can have up to ten steps in each 
autosequence. 
Making an Autosequence 
Begin by accessing the custom, autosequence, IVPUMP menu as described 
above. This time, select: 
MAKE
F1
Next select the name of the autosequence you want to customize or create. 
Follow these steps, providing the desired information: 