How to customize, Customizing a1 – Fluke Biomedical medTester 5000C User Manual
Page 91
Biomedical Test System
Autosequences
5
5-21
Humidity Monitor
You can monitor the level of relative humidity over periods of time, print the
results, and store the results in memory. To access the monitor from the
keyboard, enter:
HUMMON
F5
TOOLBOX
F2
Selections then display as:
Select
MEMORY
if you want to store the test record. This requires both the RS-
232/Printer module for printing and the 100 Record Memory module for
storage. Select
if you want to print the test record. Select
BOTH
if you
want to store and print the record. Follow these set-up instructions:
1.
Plug the humidity adapter test leads into the
EXT INPUT
jacks on
medTester’s front panel.
2.
Power on the adapter.
The medTester will prompt you to enter the following information. Entry
information can be any combination of alpha or numeric characters. After
each entry press the
ENT
key:
1.
OP CODE—The operator is the person doing the testing. This
code can consist of a maximum of three characters which can be
entered from the medTester keyboard. An example of an operator
code could be a person’s initials or employee number. The
previous operator’s code, if any, will appear. If the current
operator has not changed from previous use, press the
ENT
key.
2.
LOC—Type in the location of the device, 16 characters
maximum. The previous location code, if any, will appear. If this
occurs and is the desired location, press the
ENT
key.
3.
CONTROL #—Enter a control number with a maximum of 16
characters.
RUN HUMIDITY MONITOR A19 TO:
MEMORY
F1
F2
NEITHER
F4
BOTH
F3