Configuring for your spo2 simulator, Making an autosequence, For the index 2xl – Fluke Biomedical medTester 5000C User Manual
Page 247

Biomedical Test System
SPO2 Module
15
15-19
For the Index 2XL:
SPO2 Autosequence Number: A70
Autosequence Name: NELLCR
Manufacturer: Nellcor
Settling Time: 25 Beats
Heart Rate Limit:
± 4 BPM (beats per minute)
Step Number
SPO2
(in percent
SPO2 Limit
(in percent)
Heart Rate
(in BPM)
Amplitude
(in percent)
Base TLC
1 96
± 3
75 6.0 3500
2 98
± 3
230 6.0
3500
3 81
± 4
38
2.0 3500
4 81
± 4
112 6.0
3500
5 81
± 4
201 6.0
3500
6 92
± 3
75 6.0 3500
7 82
± 4
75 6.0 3500
8 72
± 5
75 6.0 3500
9 98
± 3
30 6.0 3500
Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 6.0%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
Probe Test
SPO2 Autosequence Number: A71
Autosequence Name: OHMNOV
Manufacturer: Ohmeda & Nova
Settling Time: 25 Beats
Heart Rate Limit:
± 4 BPM (beats per minute)
Step Number
SPO2
(in percent
SPO2 Limit
(in percent)
Heart Rate
(in BPM)
Amplitude
(in percent)
Base TLC
1 96
± 3
75 6.0 3520
2 98
± 3
230 6.0
3520
3 92
± 3
75 6.0 3520
4 82
± 4
75 6.0 3520
5 72
± 5
75 6.0 3520
6 98
± 3
45 6.0 3520
Starting Alarm Value for SPO2: 96%
Starting Alarm Value for Rate: 75 BPM
Starting Alarm Value for Amplitude: 6.0%
SPO2 Alarm Test: Hi–100% Lo–85%
Rate Alarm Test: Hi–140 BPM Lo–50 BPM
Asystole Test
Amplitude Test
Probe Test