Output[]:sync:polarity – RIGOL DG1000Z Series User Manual
Page 72

RIGOL
2-60
DG1000Z Programming Guide
:OUTPut[
Syntax
:OUTPut[
:OUTPut[
Description
Set the output polarity of the sync signal on the [Mod/Trig/FSK/Sync] connector
of the specified channel at the rear panel to normal (POSitive) or inverted
(NEGative).
Query the output polarity of the sync signal on the [Mod/Trig/FSK/Sync]
connector of the specified channel at the rear panel.
Parameter
Name
Type
Range
Default
[
Discrete
1|2
1
{POSitive|NEGative}
Discrete
POSitive|NEGative
POSitive
Explanation
When [
CH1 by default.
The output polarity of the sync signal refers to that the sync signal on the
[Mod/Trig/FSK/Sync] connector of the channel at the rear panel is normal
(POSitive) output or inverted (NEGative) output. In the normal mode, the
instrument outputs the sync signal normally and in the inverted mode, the sync
signal is inverted and then outputted.
After the waveform is inverted (
), the sync signal related
to the waveform will not be inverted.
Return Format
The query returns POS or NEG.
Example
:OUTP1:SYNC:POL POS /*Set the output polarity of the sync signal on the
[Mod/Trig/FSK/Sync] connector of CH1 at the rear
panel to normal */
:OUTP1:SYNC:POL? /*Query the output polarity of the sync signal on the
[Mod/Trig/FSK/Sync] connector of CH1 at the rear
panel and the query returns POS*/
Related
Command