Coupling:frequency:deviation – RIGOL DG1000Z Series User Manual
Page 26

RIGOL
2-14
DG1000Z Programming Guide
:COUPling:FREQuency:DEViation
Syntax
:COUPling:FREQuency:DEViation
:COUPling:FREQuency:DEViation?
Description
Set the frequency deviation in the frequency coupling.
Query the frequency deviation in the frequency coupling.
Parameter
Name
Type
Range
Default
Real
-59.999 999 999 999MHz to
59.999 999 999 999MHz
0Hz
Explanation
Select the desired frequency coupling mode (
) and
set the corresponding frequency deviation or frequency ratio
) before enabling the frequency coupling function
). You cannot set the frequency coupling mode
and frequency deviation/ratio after the frequency coupling function is enabled.
When the frequency coupling function is disabled, if the current frequency
coupling mode is frequency deviation, send this command can set the frequency
deviation; if the current frequency coupling mode is frequency ratio, send this
command can set the frequency coupling mode to frequency deviation and set
the frequency deviation.
[:SOURce[
set or query the frequency deviation in the frequency coupling.
Return Format
The query returns the frequency deviation in scientific notation. The return value
contains 7 effective digits, for example, 1.000000E+02 (the frequency deviation is
100Hz).
Example
:COUP:FREQ:DEV 100 /*Set the frequency deviation in the frequency coupling to
100Hz*/
:COUP:FREQ:DEV? /*Query the frequency deviation in the frequency coupling
and the query returns 1.000000E+02*/
Related
Commands
:COUPling:FREQuency:MODE
:COUPling:FREQuency:RATio
:COUPling:FREQuency[:STATe]
[:SOURce[