[:SOURce[]]:FREQuency:COUPle[:STATe]

Syntax

[:SOURce[]]:FREQuency:COUPle[:STATe] {ON|1|OFF|0}

[:SOURce[]]:FREQuency:COUPle[:STATe]?

Description

Turn on or off the frequency coupling function.
Query the on/off status the frequency coupling function.

Parameter

Name

Type

Range

Default

[]

Discrete

1|2

1

{ON|1|OFF|0}

Bool

ON|1|OFF|0

OFF

Explanation

 When the frequency coupling function is turned off, you can select the frequency

coupling mode and set the corresponding frequency deviation or ratio. After the

frequency coupling function is turned on, CH1 and CH2 are the reference source

of each other; when the frequency of one channel (this channel is the reference

source) is changed, the frequency of the other channel will be adjusted

automatically and the channel will always keep the specified frequency deviation

or ratio with the reference channel.

 Please select the desired frequency coupling mode

(

[:SOURce[]]:FREQuency:COUPle:MODE

) and set the corresponding

frequency deviation (

[:SOURce[]]:FREQuency:COUPle:OFFSet

) or frequency

ration (

[:SOURce[]]:FREQuency:COUPle:RATio

) before enabling the

frequency coupling function. You cannot set the frequency coupling mode and

the frequency deviation/ratio after the frequency coupling function is enabled.

 You can also send the

:COUPling:FREQuency[:STATe]

command to set or query

the status of the frequency coupling function.

Return Format

The query returns ON or OFF.

Example

:FREQ:COUP ON /*Turn on the frequency coupling function */
:FREQ:COUP? /*Query the on/off status the frequency coupling function and

the query returns ON*/

Related

Commands

:COUPling:FREQuency[:STATe]

[:SOURce[]]:FREQuency:COUPle:MODE
[:SOURce[]]:FREQuency:COUPle:OFFSet