Source[]]:voltage[:level][:immediate]:offset – RIGOL DG1000Z Series User Manual
Page 196

RIGOL
2-184
DG1000Z Programming Guide
[:SOURce[
Syntax
[:SOURce[
{
[:SOURce[
Description
Set the waveform offset voltage of the specified channel.
Query the waveform offset voltage of the specified channel.
Parameter
Name
Type
Range
Default
[
Discrete
1|2
1
Real
Refer to the “Explanation”
0V
DC
Explanation
When [:SOURce[
parameters of CH1 by default.
The range of the DC offset voltage is limited by the “Impedance”
), “Frequency”
) and “Amplitude”
[:SOURce[
) settings. If the
offset in the command is greater than the corresponding offset upper limit or
lower than the corresponding offset lower limit, the system will set the waveform
offset of the specified channel to the offset upper limit or lower limit.
The current DC offset voltage of the instrument is the default value or the offset
specified formerly. When the instrument configuration (such as the impedance)
is changed, the instrument still uses the current offset if the offset is valid; the
instrument displays prompt message and set the offset to the offset upper limit
of the new configuration automatically if the current offset is invalid.
Return Format
The query returns the waveform offset voltage in scientific notation with 7 effective
digits, for example, 1.000000E+00 (the waveform offset voltage is 1V
DC
).
Example
:SOUR1:VOLT:OFFS 1 /*Set the offset voltage of CH1 to 1V
DC
*/
:SOUR1:VOLT:OFFS? /*Query the offset voltage of CH1 and the query returns
1.000000E+00*/
Related
Commands
[:SOURce[