System:communicate:lan:ipaddress, System:communicate:lan:mac – RIGOL DG1000Z Series User Manual
Page 206

RIGOL
2-194
DG1000Z Programming Guide
:SYSTem:COMMunicate:LAN:IPADdress
Syntax
:SYSTem:COMMunicate:LAN:IPADdress
:SYSTem:COMMunicate:LAN:IPADdress?
Description
Set the IP address.
Query the IP address.
Parameter
Name
Type
Range
Default
ASCII string
Refer to the “Explanation”
None
Explanation
This command is only valid when the ManualIP configuration mode is turned
on (
:SYSTem:COMMunicate:LAN:STATic[:STATe]
).
The format of
from 1 to 223 (except 127) and the other three nnn range from 0 to 255. It is
recommended that users acquire an available address from their network
administrator.
After sending this comammd, the new settings will only take effect when
command is executed to apply the
current network parameters.
Return Format
The query returns a string, for example, 192.168.1.88.
Example
:SYST:COMM:LAN:IPAD 192.168.1.88 /*Set the IP address to 192.168.1.88*/
:SYST:COMM:LAN:IPAD? /*Query the IP address and the query
returns 192.168.1.88*/
Related
Commands
:SYSTem:COMMunicate:LAN:STATic[:STATe]
:SYSTem:COMMunicate:LAN:MAC?
Syntax
:SYSTem:COMMunicate:LAN:MAC?
Description
Query the MAC address of the instrument.
Explanation
MAC (Media Access Control) address, also called hardware address, is used for
defining the position of the network device. For a signal generator, the MAC address
is always unique. It is always used to identify the instrument while assigning IP
address for instrument. MAC address (48 bits, namely 6 bytes) is usually expressed
in hexadecimal format, such as 00-14-0E-42-12-CF.
Return Format
The query returns a string, for example, 00-14-0E-42-12-CF.