Output[]:polarity, Output[][:state – RIGOL DG1000Z Series User Manual
Page 70

RIGOL
2-58
DG1000Z Programming Guide
:OUTPut[
Syntax
:OUTPut[
:OUTPut[
Description
Set the output polarity of the specified channel to normal (NORMal) or inverted
(INVerted).
Query the output polarity of the specified channel.
Parameter
Name
Type
Range
Default
[
Discrete
1|2
1
{NORMal|INVerted}
Discrete
NORMal|INVerted
NORMal
Explanation
When [
of CH1 by default.
The output polarity of the channel refers to that the signal on the output
connector of the channel is normal (NORMal) output or inverted (INVerted)
output. In the normal mode, the instrument outputs normal waveform and in
the inverted mode, the waveform is inverted and then outputted.
The waveform is inverted relative to the offset voltage. After the waveform is
inverted, the offset voltage remains unchanged and the sync signal related to
the waveform is not inverted.
Return Format
The query returns NORMAL or INVERTED.
Example
:OUTP1:POL NORM /*Set the output polarity of CH1 to normal*/
:OUTP1:POL? /*Query the output polarity of CH1 and the query returns
NORMAL*/
:OUTPut[
Syntax
:OUTPut[
:OUTPut[
Description
Turn on or off the output of the specified channel.
Query the output status of the specified channel.
Parameter
Name
Type
Range
Default
[
Discrete
1|2
1
{ON|1|OFF|0}
Bool
ON|1|OFF|0
OFF
Explanation
When [
CH1 by default.
Return Format
The query returns ON or OFF.
Example
:OUTP1? /*Query the output status of CH1 and the query returns OFF*/
:OUTP1 ON /*Turn on the output of CH1*/
:OUTP1? /*Query the output status of CH1 and the query returns ON*/