RIGOL DG1000Z Series User Manual
Page 46

RIGOL
2-34
DG1000Z Programming Guide
*RST
Syntax
*RST
Description
Restore the instrument to its factory state.
Explanation
Restore the instrument to its factory state (please refer to “Appendix B:
Factory Setting”) and it is not affected by the
command.
This command will stop the sweep or burst in progress in an abnormal way
and the screen display will be turned on if it is previously turned off (using
Related
Commands
*SAV
Syntax
*SAV
{USER1|USER2|USER3|USER4|USER5|USER6|USER7|USER8|USER9|USER10|
ARB1|ARB2|ARB3|ARB4|ARB5|ARB6|ARB7|ARB8|ARB9|ARB10}
Description
Store the current instrument state (USER) or arbitrary waveform data (ARB) in the
specified storage location in the internal non-volatile memory with the default
name.
Parameter
Name
Type
Range
Default
{USER1|USER2|USER3|
USER4|USER5|USER6|
USER7|USER8|USER9|
USER10|ARB1|ARB2|ARB3|
ARB4|ARB5|ARB6|ARB7|
ARB8|ARB9|ARB10}
Discrete
USER1|USER2|USER3|
USER4|USER5|USER6|
USER7|USER8|USER9|
USER10|ARB1|ARB2|
ARB3|ARB4|ARB5|ARB6|
ARB7|ARB8|ARB9|ARB10
None
Explanation
The internal memory of the instrument provides 10 storage locations
(numbered 1 to 10) for the instrument state and arbitrary waveform data
respectively. The default sate file name is Scpin.RSF and the default arbitrary
waveform file name is Scpin.RAF; wherein, n corresponds to the number of the
storage location.
If the specified storage location already contains a file, this command will
store the current instrument state or arbitrary waveform data in the specified
storage location and directly overwrite the original file. If the original state file
in the specified storage location is locked (
), this
command is invalid (do not overwrite the original file).
For the introductions of the state file and arbitrary waveform file, please refer
to the “Explanation” under the
command to recall the state file and arbitrary
waveform file stored in the internal non-volatile memory of the instrument.
Example
*SAV USER1 /*Store the current instrument state in storage location 1 in the
internal non-volatile memory of the instrument with the filename
Scpi1.RSF*/
Related