RIGOL DG1000Z Series User Manual
Page 47

RIGOL
DG1000Z Programming Guide
2-35
*SRE
Syntax
*SRE
*SRE?
Description
Enable the bits in the status byte register to generate service request.
Query the bits enabled in the status byte register.
Parameter
Name
Type
Range
Default
Integer
Refer to the “Explanation”
None
Explanation
the bits enabled in the status byte register. The bits selected are accumulated
on bit6 (main accumulation bit) of the status byte register and service request
will be generated if any of the bits selected changes from 0 to 1.
When
register of the status byte register.
If you have configured the instrument using the
*PSC 1
command, the enable
register of the status byte register will be cleared at the next power-on of the
instrument; if you have configured the instrument using the
*PSC 0
command,
the enable register of the status byte register will not be cleared at the next
power-on of the instrument.
Return Format
The query returns a decimal value corresponding to the sum of the binary weights
of the bits enabled in the status byte register.
Related
Command
*STB?
Syntax
*STB?
Description
Query the status register of the status byte register.
Explanation
This command cannot clear the service request. Bit6 (main accumulation bit) of the
status byte register will not be cleared as long as the condition that generates the
service request is still in effect.
Return Format
The query returns a decimal value corresponding to the sum of the binary weights
of all the bits in the status register of the status byte register.