Source[]][:mod]:pm:source – RIGOL DG1000Z Series User Manual
Page 151

RIGOL
DG1000Z Programming Guide
2-139
[:SOURce[
Syntax
[:SOURce[
[:SOURce[
Description
Set the PM modulation source of the specified channel to internal (INTernal) or
external (EXTernal) modulation source.
Query the PM modulation source of the specified channel.
Parameter
Name
Type
Range
Default
[
Discrete
1|2
1
{INTernal|EXTernal}
Discrete
INTernal|EXTernal
INTernal
Explanation
DG1000Z can receive the modulation waveform from the internal or external
modulation source.
When internal modulation source is selected, the modulation waveform can be
SINusoid, SQUare, TRIangle, RAMP, NRAMp, NOISe or USER and the default is
SINusoid. NOISe can be used as modulation waveform but cannot be used as
carrier waveform.
When external modulation source is selected, the signal generator receives the
external modulating signal from the corresponding [Mod/Trig/FSK/Sync]
connector at the rear panel. At this point, the phase deviation of the
modulated waveform is controlled by the ±5 V signal level of the connector.
For example, if the phase deviation is set to 180°, +5V signal level
corresponds to a 180º phase variation. The lower the external signal level, the
less deviation would be generated.
When [:SOURce[
parameters of CH1 by default.
Return Format
The query returns INT or EXT.
Example
:SOUR1:PM:SOUR EXT /*Set the PM modulation source of CH1 to external
modulation source*/
:SOUR1:PM:SOUR? /*Query the FM modulation source of CH1 and the
query returns EXT*/