Source:sweep commands, Source[]]:sweep:htime:start, Source:sweep commands -162 – RIGOL DG1000Z Series User Manual
Page 174

RIGOL
2-162
DG1000Z Programming Guide
:SOURce:SWEep Commands
[:SOURce[
Syntax
[:SOURce[
[:SOURce[
Description
Set the start hold of the sweep function of the specified channel.
Query the start hold of the sweep function of the specified channel.
Parameter
Name
Type
Range
Default
[
Discrete
1|2
1
Real
0s to 500s
0s
Explanation
Start hold is the duration that the output signal outputs with the “Start
Frequency” (
) after the sweep starts. After
the start hold, the signal generator will output with varying frequency according
to the current sweep type.
When [:SOURce[
parameters of CH1 by default.
After the “Start Hold” is modified, the signal generator will output the sweep
waveform from the specified “Start Frequency”.
Return Format
The query returns the start hold in scientific notation with 7 effective digits, for
example, 1.000000E+00 (the start hold is 1s).
Example
:SOUR1:SWE:HTIM:STAR 1 /*Set the start hold of the sweep function of CH1 to
1s*/
:SOUR1:SWE:HTIM:STAR? /*Query the start hold of the sweep function of CH1
and the query returns 1.000000E+00*/
Related
Command