:SYSTem:ROSCillator:SOURce
Syntax
:SYSTem:ROSCillator:SOURce {INTernal|EXTernal}
:SYSTem:ROSCillator:SOURce?
Description
Set the system clock source to internal source (INTernal) or external source
(EXTernal).
Query the system clock source type.
Parameter
Name
Type
Range
Default
{INTernal|EXTernal}
Discrete
INTernal|EXTernal
INTernal
Explanation
DG1000Z provides an internal 10MHz clock source and accepts external clock
source from the [10MHz In/Out] connector at the rear panel. It can also
output clock source from the [10MHz In/Out] connector for other devices.
If external clock source is selected, the instrument will detect whether a valid
external clock signal is inputted from the [10MHz Out/In] connector at the
rear panel. If no valid clock source is detected, the corresponding prompt
message would be displayed and the clock source would be switched to
internal.
You can synchronize two or more instruments by setting the clock source.
Synchronization between two instruments:
Connect the output of the [10MHz In/Out] connector of instrument A (set
the clock source to “Internal”) to the [10MHz In/Out] connector of
instrument B (set the clock source to “External”) and set the output
frequencies of instruments A and B as a same value to realize synchronization
between the two instruments.
Synchronization among multiple instruments:
Divide the 10MHz clock source of a instrument (set the clock source to
“Internal”) into multiple instruments, and then connect them to the [10MHz
In/Out] connectors of other generators (set the clock source to “External”)
respectively, and finally set the output frequencies of all the instruments as a
same value to realize synchronization among the multiple instruments.
Return Format
The query returns INT or EXT.
Example
:SYST:ROSC:SOUR INT /*Set the system clock source to internal*/
:SYST:ROSC:SOUR? /*Query the system clock source and the query returns
INT*/
:SYSTem:SECurity:IMMediate
Syntax
:SYSTem:SECurity:IMMediate
Description
Sanitizes all the user-accessible instrument memory information.
Explanation
This command sanitizes all the user-defined state information, arbitrary waveforms
and I/O settings (such as the IP address) as well as restores the instrument settings
to the factory values.