Source[]]:sweep:state – RIGOL DG1000Z Series User Manual
Page 178

RIGOL
2-166
DG1000Z Programming Guide
[:SOURce[
Syntax
[:SOURce[
[:SOURce[
Description
Turn on or off the sweep function of the specified channel.
Query the on/off status of the sweep function of the specified channel.
Parameter
Name
Type
Range
Default
[
Discrete
1|2
1
{ON|1|OFF|0}
Bool
ON|1|OFF|0
OFF
Explanation
DG1000Z can output sweep waveform from a single channel or both of the two
channels at the same time. In the sweep mode, the signal generator outputs
from the start frequency to the stop frequency within the specified sweep time.
DG1000Z supports linear, log and step sweeps and allow users to set the “Mark”
frequency, start hold, stop hold and return time. Besides, it supports internal,
external and manual trigger sources and can generate sweep output for sine,
square, ramp and arbitrary waveform (except DC) (does not support the sweep
signals for the pulse and noise in the basic waveforms).
When [:SOURce[
parameters of CH1 by default.
If the modulation (
) or burst
) function is currently turned on, it will be
turned off automatically when the sweep function is turned on.
If the harmonic function (
) is currently
turned on, the sweep function cannot be turned on.
Return Format
The query returns ON or OFF.
Example
:SOUR1:SWE:STAT ON /*Turn on the sweep function of CH1*/
:SOUR1:SWE:STAT? /*Query the on/off status of the sweep function of CH1
and the query returns ON*/
Related
Commands
[:SOURce[