Symbol description, Parameter type, Symbol description -5 – RIGOL DG1000Z Series User Manual
Page 11: Parameter type -5

Chapter 1 Programming Overview
RIGOL
DG1000Z Programming Guide
1-5
Symbol Description
The following symbols are not the content of the SCPI commands and will not be sent with the commands.
They are usually used to describe the parameters in the commands.
1. Braces {}
Multiple optional parameters are enclosed in the braces and one of the parameters must be selected
when sending the command. For example, :DISPlay:MODE {DPV|DGV|SV}.
2. Vertical Bar |
The vertical bar is used to separate multiple parameters and one of the parameters must be selected
when sending the command. For example, :DISPlay:MODE {DPV|DGV|SV}.
3. Square Brackets []
The content (command keyword or parameter) in the square brackets can be omitted. If the parameter
is omitted, the instrument will set the parameter to its default. For example, for
the :COUNter:STATIstics[:STATe]? command, send any of the following two commands can achieve
the same effect.
:COUNter:STATIstics?
:COUNter:STATIstics:STATe?
4. Triangle Brackets <>
The parameter enclosed in the triangle brackets must be replaced by an effective value. For example,
sending the :COUNter:LEVEl
Parameter Type
The parameters of the commands introduced in this manual contain the following 5 types: bool, integer,
real number, discrete and ASCII string.
1 Bool
The parameter could be ON (1) or OFF (0). For example, :COUNter:HF {ON|1|OFF|0}.
2 Integer
Unless otherwise noted, the parameter can be any integer within the effective value range. Note that
do not set the parameter to a decimal; otherwise, errors will occur. For example, in
the :DISPlay:BRIGhtness
3 Real Number
Unless otherwise noted, the parameter can be any real number within the effective value range.
For example, the range of
4 Discrete
The parameter could only be one of the specified values or characters. For example, in
the :DISPlay:MODE {DPV|DGV|SV} command, the parameter can only be DPV, DGV or SV.
5 ASCII String
The parameter should be the combinations of ASCII characters. For example, in
the :MMEMory:LOAD:STATe
loaded under the current directory of the external memory and can include English characters and
numbers.
Besides, you can replace the parameters in many commands with MINimum or MAXimum to set the
parameters to their minimum or maximum value. For example, MINimum and MAXimum in
the :DISPlay:BRIGhtness {
the minimum or maximum.