Source[]]:sweep:time, Source[]]:sweep:trigger[:immediate – RIGOL DG1000Z Series User Manual
Page 180

RIGOL
2-168
DG1000Z Programming Guide
[:SOURce[
Syntax
[:SOURce[
[:SOURce[
Description
Set the sweep time of the specified channel.
Query the sweep time of the specified channel.
Parameter
Name
Type
Range
Default
[
Discrete
1|2
1
Real
1ms to 500s
1s
Explanation
The sweep time refers to the time required for the instrument to sweep from
the “Start Frequency” (
) to the “Stop
Frequency” (
).
When [:SOURce[
parameters of CH1 by default.
After the “Sweep Time” is modified, the signal generator will output the sweep
waveform from the specified “Start Frequency”.
Return Format
The query returns the sweep time in scientific notation with 7 effective digits, for
example, 5.000000E+00 (the sweep time is 5s).
Example
:SOUR1:SWE:TIME 5 /*Set the sweep time of CH1 to 5s*/
:SOUR1:SWE:TIME? /*Query the sweep time of CH1 and the query returns
5.000000E+00*/
Related
Commands
[:SOURce[
Syntax
[:SOURce[
Description
Trigger a sweep immediately in the specified channel.
Parameter
Name
Type
Range
Default
[
Discrete
1|2
1
Explanation
This command is only applicable to manual trigger
[:SOURce[
) and is only valid when the output
of the corresponding channel is turned on.
When [:SOURce[
parameters of CH1 by default.
Example
:SOUR1:SWE:TRIG /*Trigger a sweep immediately in CH1*/
Related
Commands
[:SOURce[