2-5
:COUNter:HF
Syntax
:COUNter:HF {ON|1|OFF|0}
:COUNter:HF?
Description
Enable or disable the high-frequency rejection function of the frequency counter.
Query the on/off status of the high-frequency rejection function of the frequency
counter.
Parameter
Name
Type
Range
Default
{ON|1|OFF|0}
Bool
ON|1|OFF|0
OFF
Explanation
Enable the high-frequency rejection when measuring low-frequency signal with
lower than 250kHz frequency to filter out the high-frequency noise and improve
the measurement accuracy. Disable the high-frequency rejection when measuring
high-frequency signal with greater than 250kHz frequency; at this point, the
maximum input frequency is 200MHz.
Return Format
The query returns ON or OFF.
Example
:COUN:HF ON /*Enable the high-frequency rejection function of the frequency
counter*/
:COUN:HF? /*Query the on/off status of the high-frequency rejection
function of the frequency counter and the query returns ON*/
:COUNter:LEVEl
Syntax
:COUNter:LEVEl {
:COUNter:LEVEl? [MINimum|MAXimum]
Description
Set the trigger level of the frequency counter.
Query the trigger level of the frequency counter.
Parameter
Name
Type
Range
Default
Real
-2.5V to 2.5V
0V
Explanation
The frequency counter starts measuring when the input signal reaches the
specified trigger level.
The minimum resolution is 6mV.
Return Format
The query returns the trigger level in scientific notation. The return value contains 7
effective digits, for example, 1.500000E+00 (the trigger level is 1.5V).
Example
:COUN:LEVE 1.5 /*Set the trigger level of the frequency counter to 1.5V*/
:COUN:LEVE? /*Query the trigger level of the frequency counter and the
query returns 1.500000E+00*/