Status indicators – Rockwell Automation 1441-PEN25-Z Enpac 2500 Data Collector User Manual
Page 25
Rockwell Automation Publication GMSI10-UM002D-EN-E - August 2012
25
The Enpac 2500 Data Collector
Chapter 1
Status Indicators
The Enpac 2500 has three status indicators located in the upper-right side of the
Enpac nameplate.
Figure 4 - Status Indicators
Red
Amber
Green
Table 4 - Status Indicators
Status Indicators
Operation
State
Definition
Red
Charging Battery
Flashing
When first connected to the power adapter, the Enpac 2500 tests the
condition of the main battery. Within 30 seconds, the status indicators
should go to a solid state. If the status indicators continues to flash, the unit
has diagnosed a fault in the main battery and it will not charge the battery.
The problem may be that the cell temperature is too high or there is a fault
with the battery.
Solid
The main battery is OK. The power adapter is connected to the Enpac 2500
and charging the main battery. The battery will be maintained by a trickle
charge from the power adapter for as long as it remains attached to the unit.
Data Collection
Solid
• An alarm condition has been set.
• Input signal is out of range.
• ICP Fail is active.
• The power adapter is connected to the Enpac 2500.
Amber
Data Collection
Solid
Acquisition sub-system is settling.
Green
Data Collection
Solid
Input signal is stable and data is ready to be stored.