Rockwell Automation 1441-PEN25-Z Enpac 2500 Data Collector User Manual

Page 144

background image

144

Rockwell Automation Publication GMSI10-UM002D-EN-E - August 2012

Chapter 4

Collecting and Reviewing Data

Create a User-defined Point

Follow these steps to create a user-defined point.

1.

Select Data Collection from the Main Menu and press the Enter key.

The Data Collection screen appears.

2.

Select the route and press Enter.

The Enpac 2500 data collector always associates Offroute measurements
with a list.

3.

Press F2 (Offroute) to collect an Offroute measurement.

gSE

Performs a standard vibration Acceleration measurement and collects waveform data.
These are the settings for the measurement parameters.

Table 36 - Types of Offroute Measurements

Type

Description

Parameter

Value

Units

Enveloped Acceleration

Display Format

Spectrum

Band High-pass frequency

gSE BPF 5 kHz

Fmax

10kHz

Lines/Samples

400

Averages

4

Input Channel

X

Overlap

50%

Ext. Trigger Slope

Positive

Trigger Level

2 V

Bin Zeroing

On

Type

Spectral