Rockwell Automation 1441-PEN25-Z Enpac 2500 Data Collector User Manual
Page 167

Rockwell Automation Publication GMSI10-UM002D-EN-E - August 2012
167
Multi-channel Measurements Chapter 6
Use the descriptions in the table below to help you configure the options in
the Offroute dialog box.
When you are finished setting up the parameters, you can collect the
Offroute measurement.
Collect Orbit Measurements on page 168
for more information.
TIP
The Units, Transducer, Filter, and Sensitivity apply to both channels.
The two input channels are X and Y.
Table 39 - Offroute Orbit Parameters
Parameter
Description
Values/Comments
Units
Sets the measurement units for the transducer.
Disp (mil)
Disp (μm)
Transducer
The transducer used to collect this measurement.
Prox Probe
Accel
Sensitivity
The sensitivity of the transducer.
The sensitivity value is included with the transducers
documentation or it may be imprinted on the side of the
transducer.
200 mV/EU (default)
Trigger
Determines whether the trigger is on or off.
When Trigger is set to Off the rpm value must be entered
during data collection.
On
Off
Ext Trig Slope
The input signal slope to be used for externally triggering the
measurement
+ve
Defines the trigger event to occur on a rising edge.
-ve
Defines the trigger event to occur on a falling edge.
+ve
-ve
Level
The amplitude level for externally triggering the
measurement.
The level is a percentage of the trigger full scale range.
2.0 (default)