Multi-channel Measurements Chapter 6

Use the descriptions in the table below to help you configure the options in
the Offroute dialog box.

When you are finished setting up the parameters, you can collect the
Offroute measurement.

See

Collect Orbit Measurements on page 168

for more information.

TIP

The Units, Transducer, Filter, and Sensitivity apply to both channels.
The two input channels are X and Y.

Table 39 - Offroute Orbit Parameters

Parameter

Description

Values/Comments

Units

Sets the measurement units for the transducer.

Disp (mil)
Disp (μm)

Transducer

The transducer used to collect this measurement.

Prox Probe
Accel

Sensitivity

The sensitivity of the transducer.
The sensitivity value is included with the transducers
documentation or it may be imprinted on the side of the
transducer.

200 mV/EU (default)

Trigger

Determines whether the trigger is on or off.
When Trigger is set to Off the rpm value must be entered
during data collection.

On
Off

Ext Trig Slope

The input signal slope to be used for externally triggering the
measurement
+ve
Defines the trigger event to occur on a rising edge.
-ve
Defines the trigger event to occur on a falling edge.

+ve
-ve

Level

The amplitude level for externally triggering the
measurement.
The level is a percentage of the trigger full scale range.

2.0 (default)