Adding the correction weight and taking a, Residual measurement – Rockwell Automation 1441-PEN25-Z Enpac 2500 Data Collector User Manual
Page 218

218
Rockwell Automation Publication GMSI10-UM002D-EN-E - August 2012
Chapter 8
Balancing Application
The following message appears asking if the trial weight is still attached.
5.
Press F2 (Yes) if the weight is still attached to the rotor. Press F3 (No) if the
weight has been removed from the rotor. Note that it is good practice to
remove the trial weight.
6.
Shut down the machine.
7.
Continue the balancing procedure in the next section.
Adding the Correction Weight and Taking a Residual Measurement
The Enpac 2500 calculates and displays a recommended balancing weight in the
Correction Weight - Solution screen.
You can split the correction weight into two distinct angles, as well as modify the
radius of the correction weight. Additionally, the Enpac 2500 calculates the
amount of weight that might be removed if weights are already attached to the
rotor.
IMPORTANT
Anytime during the initial run, you can do the following:
• Press F4 (Esc) to exit the balancing program.
• Press the 0 (Shift) key and F1 (Go to) simultaneously to go back to a
previous step in the balance run.
Refer to Moving Around in a Balance Run on page 233
for details.
• Press F2 (Summary) to display the results in the Vibration Summary
Table.
Refer to Reviewing Data on page 233
for details.