Channel Control

Background

External channel control is used with the Channel
Control connectors. Each channel control connector
can activate each load, monitor voltage and current
and has an external voltage reference input. The
voltage and current monitors output 0~100% of the
rated current/voltage as a voltage of 0~10V.

AWG 24

STRIP GAUGE

10.0 mm

1

CH CONT 1

1

CH CONT 2

1

CH CONT 3

1

CH CONT 4

1

CH CONT 5

1

CH CONT 6

1

CH CONT 7

1

CH CONT 8

SER. NO. LB

2

1

FRAME CONT

GO / NG OUTPUT

GPIB

RS232C

DISCONNECT POWER CORD
BEFORE REPLACING FUSE

50/60 Hz

180 VA MAX

AC

250V

T 3.15A

REPLACE FUSE

115V
230V

FUSE RATING

AS SPECIFIED

WARNING

TO AVOID ELECTRIC SHOCK THE POWER CORD PROTECTIVE

DO NOT REMOVE COVERS. REFER SERVICING TO QUALIFIED PERSONNEL.

250V FUSE OF THE SPECIFIED TYPE AND RATING.

FOR CONTINUED FIRE PROTECTION. REPLACE FUSE ONLY WITH

NO OPERATOR SERVICEABLE COMPONENTS INSIDE.

GROUNDING CONDUCTOR MUST BE CONNECTED TO GROUND.

External Voltage
Reference

A voltage reference of 0-10V is used to represent 0-
100% of the rating voltage/current of a load
module. As seen below the external voltage
reference and the rating voltage/current have a
linear relationship. By varying the reference voltage
between 0~10V the voltage/current setting will be
changed accordingly.