Channel control, Getting started – GW Instek PEL-2000 Series User Manual User Manual
Page 85
GETTING
STARTED
85
Channel Control
Background
External channel control is used with the Channel
Control connectors. Each channel control connector
can activate each load, monitor voltage and current
and has an external voltage reference input. The
voltage and current monitors output 0~100% of the
rated current/voltage as a voltage of 0~10V.
AWG 24
STRIP GAUGE
10.0 mm
1
CH CONT 1
1
CH CONT 2
1
CH CONT 3
1
CH CONT 4
1
CH CONT 5
1
CH CONT 6
1
CH CONT 7
1
CH CONT 8
SER. NO. LB
2
1
FRAME CONT
GO / NG OUTPUT
GPIB
RS232C
DISCONNECT POWER CORD
BEFORE REPLACING FUSE
50/60 Hz
180 VA MAX
AC
250V
T 3.15A
REPLACE FUSE
115V
230V
FUSE RATING
AS SPECIFIED
WARNING
TO AVOID ELECTRIC SHOCK THE POWER CORD PROTECTIVE
DO NOT REMOVE COVERS. REFER SERVICING TO QUALIFIED PERSONNEL.
250V FUSE OF THE SPECIFIED TYPE AND RATING.
FOR CONTINUED FIRE PROTECTION. REPLACE FUSE ONLY WITH
NO OPERATOR SERVICEABLE COMPONENTS INSIDE.
GROUNDING CONDUCTOR MUST BE CONNECTED TO GROUND.
External Voltage
Reference
A voltage reference of 0-10V is used to represent 0-
100% of the rating voltage/current of a load
module. As seen below the external voltage
reference and the rating voltage/current have a
linear relationship. By varying the reference voltage
between 0~10V the voltage/current setting will be
changed accordingly.