OPERATION

171

3. Press the Selector knob to edit

the selected setting, turn to
change the setting from Panel to
External.

4. Press the Selector knob or Enter

to confirm selection.

Channel control is now activated. To
turn Channel Control off, CH CONT
must be set to Panel again. When
Channel Control is active, EXT will be
displayed on the side panel for the
active channel.

Note

Channel Control can only be activated on the active
channel; other channels will not be affected.

Configuring the Independent Setting

Background

The Independent setting allows a channel to be
load independent from the mainframe. What this
means is that a load module with Independent set
to ON can only load from the local load module. If
the LOAD ON/OFF key is pressed from the
mainframe, the channel with Independent set to
ON will be unaffected by the mainframe, except
when running a program.

Parameter

Independent ON/ OFF