OPERATION

177

Go/NoGo

Background

Go/NoGo mode is used to set threshold limits.
When a load is within the limit(s) it is considered
to be “Go”, when the load has exceeded the limit it
is considered to be “NoGo”.

Go/NoGo limits can be set as either absolute
values (Entry Mode set to “Value”) or as a
percentage offset from a nominal (Center) value
(Entry Mode set to “Percent”).

Go/NoGo can be used in both high and low
ranges, as well as CC, CV, and CR modes. The
Go/NoGo status can be read using the rear
Go/NoGo output.

A delay time can also be imposed for up to 1
second.

Note

Any Go/NoGo configuration only applies to the
current channel with the same mode and range.

Parameters

Value Percent

CC Mode
CR Mode

High: V
Low: V

High: %
Low: %
Center: V

CV Mode

High: A
Low: A

High: %
Low: %
Center: A

Delay Time 0.0~1.0 seconds

Panel Operation Ensure the menu is in the Configuration menu. See

page160.