OPERATION

151

Edit Sequence

Background

The Sequence function can be configured to create
a unique load profile to accurately simulate loads
in real time for single or multiple loads. For details
see page 74.

Each Sequence is composed of a number of points
with customizable current/resistance, slew rate
and duration times. Each sequence can be looped
an infinite amount of times. Sequences are only
applicable for CC (Static) and CR (Static) modes.

Parameters

Value 0~rating*1.02

Ω / A

SlewRate

Load module dependent

SlewRate

Duration Time 0.0001~60,000 seconds

Panel operation

1. Choose a channel and mode.

Page 119, 122

2. Press the CHAN key, F4 (Seq.

Edit) to enter the Sequence Edit
menu.

03/02/09

10 : 30

Configure

Add

Point

Loop

Delete

Point

CC

Static

High

CH1

Previous

Menu

s_edit