PEL-2000 Series User Manual

124

Panel Operation

1. Press the F3 key until Dynamic

Range mode is selected.

Note

Changing from static to dynamic mode will only affect
the current (active) channel.

Editing CC Dynamic Parameters

Background

Dynamic Constant Current Mode has two
operating current levels, slew rates and timers.

Slew rates determine the speed at which the load
will change from one level to the next.

The timers determine how long the load
module/channel will stay at level 1 or level 2.

Ensure the menu is in CC Dynamic Mode. See page
123.

Parameters

Level1 0~rating

A

Level2

0~rating A

SlewRate

Load module dependent

SlewRate

Timer1

0.025~30000.0 ms

Timer2

0.025~30000.0 ms