Operating configurations – GW Instek PEL-2000 Series User Manual User Manual

Page 81

background image

GETTING

STARTED

81

Reverse voltage
Protection

Reverse voltage protection prevents reverse voltage
damage to the PEL-2000 series up to the specified
rating. When Reverse voltage protection has been
tripped an alarm tone will sound until the reverse
voltage is removed.

For more details please see the specifications.

Under voltage
Protection

Under voltage protection will turn off the load
when the voltage drops below a set limit.

Operating Configurations

Background

There are number of operating configuration
settings. Configuration settings are for the
following:

CC Vrange, Von Voltage, Von Latch, Short Key, CH
Cont, Independent load sync, D-time and Step
resolution settings.

CC Vrange

CC Vrange (page 165) is used to set the voltage
range as High or Low for CC mode. CC voltage
range is dependent on the load module
specifications.