beautypg.com

GW Instek PEL-2000 Series User Manual User Manual

Page 257

background image

APPENDIX

257

DYNAMIC MODE

CONSTANT CURRENT MODE
T1&T2

0.025mS ~ 10mS / Res : 1uS

10mS ~ 30S / Res : 1mS

Accuracy

1uS / 1mS + 100ppm

Slew Rate

0.8 ~ 200mA/uS 0.64 ~ 160mA/uS

6.4 ~

1600mA/uS

Slew Rate

Resolution

0.8mA/uS

0.64mA/uS

6.4mA/uS

Slew Rate setting

Accuracy

±(10% + 15us)

±(10% + 15us) ±(10% + 15us)

Current

0~5A

0~4A

0~40A

Current

Resolution

1mA

1mA

10mA

Current Accuracy

±0.4% F.S.

MEASUREMENT

VOLTAGE READBACK
Range

0~16V 0~80V

0~16V

0~80V

Resolution

0.5mV 2.5mV

0.5mV

2.5mV

Accuracy

±(0.05%set + 0.05% F.S.)

CURRENT READBACK
Range

0~5A

0~4A

0~40A

Resolution

0.15625mA

0.125mA

1.25mA

Accuracy

±(0.1%set + 0.1% F.S.)

PROTECTIVE

Over Power Protection
Range

0.9~30.6W

1.25~255W

Resolution

0.15W

1.25W

Accuracy

±(2%set +
0.25%F.S)

±(2%set + 0.25%F.S)

Over Current Protection
Range

0.0625~5.1A

0.5~40.8A

Resolution

0.0125A

0.1A

Accuracy

±(2%set +
0.25%F.S)

±(2%set + 0.25%F.S)

Over Voltage Protection
Range

1~81.6V

1~81.6V

Resolution

0.2V

0.2V

Accuracy

±(2%set +
0.25%F.S)

±(2%set + 0.25%F.S)