Run Program

Background

The Program function on the PEL-2000 series
supports a total of 12 different programs at any one
time with 10 sequences to each program. Up to 12
programs can be chained together. The Program
function is able to create a number of Go/NoGo
tests.

Program
Sequence

A program sequence is simply a single load test. A
program is a battery of each of these tests run in
succession. Each sequence loads the settings for
each channel from Memory Data (Memory MXXX).
The Memory Data stores settings such as the
operating mode and range for each channel. Each
sequence loads all channels at the same time, unless
programmed otherwise. Sequences for each
channel run synchronously.

Each Sequence has a number of configuration

options that apply to all the channels equally.

Sequence Item

Parameter Range

Memory

M001~M120

Run

Auto – Skip – Manual

On-Time

0.1 ~ 60.0

Off-Time

Off – 0.1 ~ 60.0

Short-time