Set to CR Mode

Background

The PEL 2000 series load operates in three different
modes, Constant Current (CC), Constant Voltage
(CV), Constant Resistance (CR). Constant
Resistance mode will maintain a constant resistive
load, using variable current and voltage levels.

When a channel is active, the F1
key can be used to switch
between each operating mode.

Panel Operation

1. Press the CHAN button and use

the selector knob to select a
channel.

2. Press F1 until CR mode is

displayed in the display panel.

Note

Changing the operating mode will only affect the
current (active) channel. Other channels will not be
affected by any changes.

Select CR Range

Background

Constant Resistance mode can run in high and low
range. Range is dependent on the load module.

Ensure the menu is in CR Mode. See page 128.