Saving/Recalling Memory Data to USB

Background

There are two options to save or recall Memory
data to a USB flash drive:

Save Chan Current: Saves the active Channel’s
Memory data(M001~M120) into the default file
path (20XXX_XX.M).

Save Chan All: Every channel’s Memory data
(CH1 M001~120 ~ CH8 M001~M120) will be saved
into a directory (ALL00XX) as separate files for
each channel (P0X0X_C1.M ~ P0X0X_C8.M).

Recall File: Recalls the selected file to the active
channel’s Memory (MXXX). It is not possible to
update all the channels at once, only one channel at
a time can be recalled.

The file extension *.M is used for Memory data
only.

For more information about the file structures see
page 89.

Parameters

Save All Ch Directory ALL0000 ~ ALL0099

File:

P0X0X_CX.M

Save File

File: 20XXX_XX.M

Update File File: 20XXX_XX.M

Panel operation

1. Insert a USB flash drive into the

front panel USB slot.

2. Ensure the USB path has been

set.

Page 208.