beautypg.com

GW Instek AFG-2225 User Manual User Manual

Page 8

background image

AFG-2225 User Manual

8

Fuse

WARNING

Fuse type: F1A/250V.

Only qualified technicians should replace the
fuse.

To ensure fire protection, replace the fuse only
with the specified type and rating.

Disconnect the power cord and all test leads
before replacing the fuse.

Make sure the cause of fuse blowout is fixed
before replacing the fuse.

Cleaning the
function
generator

Disconnect the power cord before cleaning the
function generator.

Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid
into the function generator.

Do not use chemicals containing harsh products
such as benzene, toluene, xylene, and acetone.

Operation
Environment

Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
and avoid strong magnetic fields.

Relative Humidity: < 80%

Altitude: < 2000m

Temperature: 0°C to 40°C